dc.contributor.author | Wasunna, A | |
dc.contributor.author | Whitelaw, AG | |
dc.date.accessioned | 2013-11-05T09:12:43Z | |
dc.date.available | 2013-11-05T09:12:43Z | |
dc.date.issued | 1987-09 | |
dc.identifier.citation | Arch Dis Child. 1987 Sep;62(9):957-8 | en |
dc.identifier.uri | http://www.ncbi.nlm.nih.gov/pubmed/3674949 | |
dc.identifier.uri | http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1778566/ | |
dc.identifier.uri | http://erepository.uonbi.ac.ke:8080/xmlui/handle/123456789/58204 | |
dc.description.abstract | One hundred and twenty five measurements of arterial oxygen saturation (Stcao2) obtained with a transcutaneous pulse oximeter were compared with direct arterial oxygen tension (Pao2) in 13 preterm infants with predominantly fetal haemoglobin. Stcao2 of 86-92% corresponded to Pao2 of 5-13 kPa. Stcao2 above 92%, however, was sometimes associated with Pao2 above 13 kPa | en |
dc.language.iso | en | en |
dc.publisher | University of Nairobi, | en |
dc.title | Pulse oximetry in preterm infants | en |
dc.type | Article | en |
local.publisher | Department of Paediatrics and Neonatal Medicine | en |