dc.description.abstract | Electrochemical sensor validation en
sures that the determination procedure employed for a specific
test is suitable for its intended use. Accuracy and precision, linearity, range, limits of detection and
quantification validation parameters were determined for the Polyacrylic acid/Glassy Car
bon
Differential Pulse Anodic Stripping Voltammetric Sensor used for the simultaneous analysis of Pb(II),
Cd(II) and Co(II) in accordance with the guidelines of the ISO/IEC 17025. T
he recovery degree for
accuracy were in the range of 101% to 110% for Pb(II
), 100% to 105% for Cd(II) and 93% to 104%
for Co(II).
The precision was found to be less than 10% for ten determinations. Linear concentration
range were also investigated and found to lie in the range of 125
–
7.8μM Pb(II), 16
–
2μM Cd(II) and
2
–
0.125m
M Co(II). Limits of detection also were found to be 0.9nM Pb, 1.9mM Cd and 11.0μM Co
and limits of quantitation were 3.0nM Pb, 6.3nM Cd and 36.7μM Co. The effects of foreign substances
like Cu
2+
, SO
4
2
-
, K
+
, Na
+
, Cl
-
, NH
4
+
and O
2
were found to have no signi
ficant effect on the
electrochemical responses of the three heavy metals. These results confirms that this sensor can
provide accurate, reliable and consistent results for the determination of Pb(II), Cd(II) and Co(II)
heavy metals in diverse samples. | en_US |
dc.language.iso | en | en_US |
dc.publisher | University of Nairobi | en_US |
dc.rights | Attribution-NonCommercial-NoDerivs 3.0 United States | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/us/ | * |
dc.subject | Validation, Polyacrylic acid/Glassy carbon electrode, Lead, Cadmium, Cobalt and differential pulse anodic stripping voltammetry. | en_US |
dc.title | Validation of the polyacrylic acid/glassy carbon differential pulse anodic stripping voltammetric sensor for simultaneous analysis of lead(ii), cadmium(ii) and cobalt(ii)ions | en_US |
dc.type | Article | en_US |