Problems of indwelling Foley catheters
View/ Open
Date
1994Author
Ndirangu, K
Ngugi, M
Type
ArticleLanguage
enMetadata
Show full item recordAbstract
One hundred patients undergoing urinary bladder catheterization for various reasons were studied. 42 patients received Sewoon catheters of various sizes. 38 received Eschmann catheters of various sizes and 20 received Bard catheters of various sizes. The duration of catheterization ranged from a few hours to 21 days. Various problems associated with catheterization were recorded whenever encountered. Stuck catheter was found to occur in 15 of the 40 patients who received Sewoon catheters. There were no stuck catheters recorded for Eschmann or Bard catheters. Infections occurred in 9 out of 100 patients, 7 of whom had Sewoon catheters. Leakage was recorded in 12 of the 100 patients, 9 of whom had Sewoon catheters. Blockage of balloon occurred in 5 patients, 3 of whom had Sewoon catheters. Painful spasms occurred in 8 patients, 4 with Bard, 4 with Sewoon catheters. It is concluded that Sewoon type of Foley catheters was associated with more problems than the other types studied
URI
http://www.ncbi.nlm.nih.gov/pubmed/7925043http://erepository.uonbi.ac.ke:8080/xmlui/handle/123456789/29193
Citation
East Afr Med J. 1994 Feb;71(2):132-4Publisher
Department of Surgery, University of Nairobi, Kenya
Collections
- Faculty of Health Sciences (FHS) [10377]