Evaluation of the alkaline haematin D-575 method for haemoglobin estimation in east Africa
View/ Open
Date
1994Author
Lema, O. E
Carter, J. Y
Arube, P. A
Munafu, C. G
Wangai, M. W
Rees, P. H
Type
ArticleLanguage
enMetadata
Show full item recordAbstract
In many health facilities in east Africa, haemoglobin estimation is performed using visual colour comparison methods. Efforts to establish colorimetric methods face numerous constraints, including the unavailability of standards for quality control. In contrast, the alkaline haematin D-575 method for haemoglobin estimation is a colorimetric method that uses primary standards prepared from pure, crystalline chlorohaemin. There is no significant difference in the accuracy of the alkaline haematin D-575 method and that of the reference haemiglobincyanide method (P > 0.05), and the response of the alkaline haematin D-575 method is linear for serially diluted blood samples over the haemoglobin concentration range 19.6-3.3 g/dl (r = 0.994, y = 1.01 x - 0.3). The method has a precision of +/-0.3 g/dl (coefficient of variation = (1.8%) for whole blood, and is suitable for use with fixed-wavelength haemglobinometers (lambda = 565 nm) or with colorimeters at lambda = 580 nm. Stable quality control standards could be prepared at provincial, zonal, or reference laboratories and distributed regularly to outlying laboratories.