Show simple item record

dc.contributor.authorPratt, D. J
dc.contributor.authorPapagiannopoulos, G
dc.contributor.authorRees, P. H
dc.date.accessioned2013-06-10T06:49:05Z
dc.date.available2013-06-10T06:49:05Z
dc.date.issued1987
dc.identifier.citationJ Biomed Eng. 1987 Jan;9(1):84-7.en
dc.identifier.urihttp://hinari-gw.who.int/whalecomwww.ncbi.nlm.nih.gov/whalecom0/pubmed/3795909
dc.identifier.urihttp://erepository.uonbi.ac.ke:8080/xmlui/handle/123456789/30387
dc.description.abstractThis paper is the second part of a study into the mechanical effectiveness of a new intramedullary femoral nail (the Derby nail). The results of three point bending tests have shown that the Derby nail is stronger and stiffer than both the K and Zickel nails and offers advantages over the AO plate, despite the latter's higher load carrying ability when used on the compressively loaded bone face. Previous work by the authors has shown the Derby nail to be superior in torsional loading and it is felt that this intramedullary device offers sound clinical advantages over the other designs, this being confirmed by extensive clinical trials.en
dc.language.isoenen
dc.titleDerby intramedullary nail: further biomechanical comparisonen
dc.typeArticleen


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record