Show simple item record

dc.contributor.authorBriesen, S
dc.contributor.authorSchwering, M Schulze
dc.contributor.authorRoberts, H
dc.contributor.authorKollmann, M
dc.contributor.authorStachs, O
dc.contributor.authorBehrend, D
dc.contributor.authorSchäfer, St
dc.contributor.authorGuthoff, R
dc.date.accessioned2014-07-11T12:38:56Z
dc.date.available2014-07-11T12:38:56Z
dc.date.issued2010
dc.identifier.citationEye (2010) 24, 1279–1283en_US
dc.identifier.urihttp://hdl.handle.net/11295/72800
dc.description.abstractAims Recently, a new rebound tonometer has been introduced into the market, which might be useful for glaucoma screenings in developing countries. Disposable probes, that are potentially reusable, are recommended by the manufacturer. Our study aimed to address the question of microbial transmission risks if the probes are reused. Methods IOP measurements were obtained from 100 healthy eyes. The used probes were inoculated on broth and culture media. In addition, 10 probes were analyzed using environmental scanning electron microscopy in saturated hydrogen-steam atmosphere after usage and wipe disinfection technique with Sekusept 4% solution or Isopropanol 70%. Results No bacterial or fungal growth could be detected in any of the inoculated agar plates or broth tubes. No microorganisms, clumps of cells, or single intact epithelium cells were detected in any of the probes using environmental scanning electron microscopy. Cell debris was detected on seven probes; three probes were completely free of any residual cell elements. Conclusion Transmission of possibly infective material through reused probes is significantly less than for reusable Goldmann probes if the same sterilization protocols are applied. Re-usage of the probes appears safe and is helpful in avoiding unnecessary costs.en_US
dc.language.isoenen_US
dc.publisherUniversity of Nairobi,en_US
dc.subjectrebound tonometer; transmission; cross-infection; disinfection; glaucoma; developing countriesen_US
dc.titleMinimal cross-infection risk through Icare rebound tonometer probes: a useful tool for IOP-screenings in developing countriesen_US
dc.typeArticleen_US
dc.type.materialenen_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record